C
ALIBRATION
PROCESS
K0553 Revision D
113
[EN] English
Suggested calibration check points
Table 4-4:
Calibration check points
Ps (Calibration check points)
Pt (Calibration check points)
56.39 mbar
(~65,000 ft)**
56.39 mbar**
71.71 mbar
(~60,000 ft)*
71.71 mbar*
92.00 mbar
(~55,000ft)
92.00mbar
178.74 mbar
(~41,000ft)
178.74mbar
314.85 mbar
(~29,000ft)
314.85mbar
465.63 mbar
(~20,000ft)
465.63mbar
696.82 mbar
(~10,000ft)
696.82mbar
843.07 mbar
(~5,000ft)
843.07mbar
1013.25 mbar
(~0ft)
1013.25mbar
1128.03 mbar
(~-3,000ft)
1128.03mbar
-
-
1500mbar
-
-
1997mbar
Note: ~means approximately
a
a. *ADTS552/3/4, ** ADTS552/3/4 EALT option.