Report Formats
Report
Description
12 or 15-Lead
Report
Based on Exercise report setups, a variety of 12 or 15 Lead report
formats will print without ECG analysis when the
12 ld
key is pressed
or when 12/15 lead reports are configured in the protocol.
5 Second Rhythm
Report
This report can be chosen from the Edit Protocol application to print
at certain points during the test.
Rhythm Report
A continuous, real-time recording of raw data - 3, 6, 12 leads. Leads
for rhythm report correspond to leads on the screen.
Arrhythmia Report
Automatic documentation of arrhythmias with 2.5 seconds of raw
data prior to the ectopic beat. Leads of arrhythmia report correspond
to leads on the screen.
Recall Report
A delayed recording of raw data 10 seconds in duration. Leads of
recall report correspond to leads on the screen.
Median Report
Based on exercise setups, a Linked Median, Medians & Rhythms,
6 or 12 Lead Comparative Medians & Rhythm report will print. See
“Additional Reports” on page 174
for a description of these formats.
Exercise Final Reports
The following formats are available for the final exercise report.
Report
Description
Summary Report
One page overview of test with Resting and Max ST or peak median
morphologies. For Maximal ST Depression, report only prints when a
minimum of -.5 mm of ST depression occurs in one of the following
leads: I, II, III, aVF, V2-V6. (V1, aVR, aVL excluded. For elevation, - aVR
is excluded.)
Tabular Report
Tabular summary of test by stage including time, speed, grade,
workload, MET level, heart rate, blood pressure, RPP, and comments.
Selected Medians
Report
Records median morphologies at Baseline, Maximum ST Depression,
Peak Exercise, and Test End for 12 leads. For Maximum ST medians,
the column only prints when a minimum of -.5 mm of ST depression
occurs in one of the following leads: I, II, III, aVF, V2-V6. (V1, aVR, aVL
excluded. For elevation, - aVR is excluded.)
Trends and
Medians
Records a plot of the heart rate and blood pressure against time. Next
to these trend graphs are channels of stored median data from the
various stages of an exercise test.
Median Report
Records median morphologies for 3, 6 or 12 leads. The 3 and 6 lead
reports are configured using the 'Swedish format rhythm leads' of the
Exercise Setups (Report Leads). The 12-lead report uses the standard
12-lead set. The median storage intervals (also referred to as sample
cardiac cycles) can be configured using the Median (First and Repeat)
column of the Protocol Editor.
Trend Reports
Records plots of PVCs, heart rate, and blood pressure. Also produces
a trend report of ST level and slope vs. time. The 3-lead trend report
will use the first three leads of the ‘Swedish format rhythm leads’.
ST/HR Loops Report
A two-dimensional representation of ST Level vs. Heart Rate.
ST/HR Slope Report
Records linear regression of heart rate-adjusted slope for all leads,
plus median morphology of the lead with the highest slope.
176
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......