Recording an ECG
Continue
to close the message and continue. See
page 136
for more information on configuring Critical Values.
• The ECG report prints.
This occurs only if the report has been configured to print automatically. See
Chapter 14
, for more information.
• The
Saving ECG
window flashes on the screen.
This occurs only if a report has been set to save automatically. See
Chapter
,
, for more information.
• The ECG is transmitted to the MUSE CV system.
This occurs only if the system has been set up to transmit ECGs automatically.
See
Chapter 14
,
, for more information.
5.
To print the ECG, press the
.
6.
To save the ECG, select
Store
.
This stores the ECG either to internal memory or to an external SD card,
depending on whether the
SD Card Storage Only
option is enabled in
Miscellaneous Setup
.
NOTE:
This option is available only if the system was not configured to save the
file automatically.
After the ECG has been stored, it can be transmitted to a MUSE system. If the
system is set up to automatically transmit data, the message
Establishing
network connection
will be displayed on the screen. If the system is not
set up to automatically transmit data, you will need to manually initiate the
transmission. Refer to
Chapter 9
.
7.
Do one of the following:
• To take another ECG for the same patient, press
Same Pat
.
• To take an ECG for another patient, press
Next Pat
.
Signal Averaged ECGs
This is an optional method used for Hi-Res and PHi-Res ECGs.
1.
Select
Hi-Res
or
PHi-Res
to enable the system to record a signal averaged ECG.
NOTE:
GE Healthcare recommends a target noise level of 0.3 μV or less when
recording a PHi-Res ECG.
2.
Select
Template
to initiate the signal averaged ECG recording.
3.
Use the following steps to change the seed beat:
a.
Select
Display
.
b.
Select
SelectQRS
.
c.
Select a new seed beat.
4.
Select
Average
to average the ECG data.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
67
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......