System Setup
Setting Up Signal Averaged ECGs
To configure how the system calculates signal averages ECGs, log on to the
System
Setup
screen, select
Hi-Res
, and complete the fields described in the following table.
NOTE:
These settings are available only if you have purchased the optional
HI-Res
or
PHi-Res
functions.
Signal Averaged ECGs
Field
Description
Analysis filter
Defines the analysis filter the
Hi-Res
and
PHi-Res
options use. GE
Healthcare recommends using the
40–25Hz
filter.
Averaging target
Determines which method to use to average the target.
Target Beat Count
Determines the target heart rate. Enter a value from 1 to 999. GE
Healthcare recommends averaging to a minimum of 250 beats.
Target Noise Level
Determines the target level of noise in microvolts. Enter a value from
0.1 to 1.0. GE Healthcare recommends averaging a noise level of
0.3 microvolts.
Correlation
Threshold
Determines the degree of correlation threshold. GE Healthcare
recommends the
Very High
setting.
Final Report
Determines how many copies of the final report to print . Enter a
value from 0 to 10.
Prompt
Allows you to define custom patient questions.
Type
Allows you to set the type of response for the custom patient
question. Valid values are:
•
Numbers and letters
Allows the entry of any alphanumeric value.
•
Numbers only
Allows the entry of numeric values only.
•
Yes or No
Allows only Yes and No answers.
Setting Up Exercise Tests
This section describes the following stress test settings:
• Miscellaneous Setup
• Patient Questions
• Writer Setup
• Exercise Reports
• Screen
• Inputs and Outputs
2046275-016B
MAC
®
5500/MAC
®
5500 HD
139
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......