System Setup
Use this option if you want to use the settings to quickly configure other MAC
systems. Be sure that an SD card is inserted into the system's card slot.
• To cancel, select
Do not save setup
.
7.
Select
Main Menu
to exit
System Setup
.
Setting Up Automatic Tasks
The MAC system can be configured to perform the following tasks automatically:
• Power up into a specific function
• Preview the ECG before analysis
• Print a resting ECG report
• Print a signal averaged ECG report
• Perform ACI-TIPI Interpretation
• Store an ECG
• Transmit an ECG report
Setting Up Power Up Function
You can set up the system to start up into the function you use most often.
1.
Log into the
System Setup
function.
Refer to
“Process Overview” on page 107
.
2.
Select
Basic System
.
3.
Select
Power Up Options
.
4.
Select the function you want the system to load on startup.
For more information, refer to
“Power Up Options” on page 121
.
5.
Press
Return
.
Setting Up ECG Preview
Preview allows you to review the ECG on screen before it is printed or stored. After
the ECG is displayed, you can select
Continue
to print or store the ECG or
Cancel
to discard it.
1.
Log into the
System Setup
function.
Refer to
“Process Overview” on page 107
.
2.
Select
ECG
.
The ECG options are displayed.
3.
Select
ECG Analysis
.
The analysis options are displayed.
4.
In the
Preview before analysis
field, select
Yes
.
5.
Press
Return
.
108
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......