Exercise Stress Testing
The exercise test advances automatically through the exercise stages unless
the operator manually overrides the test.
NOTE:
When the stages in the treadmill protocol have durations other than infinite,
the exercise test advances from stage to stage automatically. However,
you can press
Exercise
(on the treadmill controller keyboard) at any time
to manually advance to the next exercise stage.
3.
Manually adjust the treadmill's speed and grade, or the ergometer's load, as
necessary.
You can adjust the speed, grade, or load in the following ways. However, once
you press these keys, the pre-programmed protocol becomes inactive: you must
manually control the speed and grade during the remainder of the exercise
and recovery phases.
• Press
Speed W +
(to increase speed) within 5 seconds of your last workload
change.
• Press
Speed W –
(to decrease speed) within 5 seconds of your last workload
change.
• Press
Grade +
(to increase grade) within 5 seconds of your last workload
change.
• Press
Grade –
(to decrease grade) within 5 seconds of your last workload
change.
The following menu items are available during the exercise phase of the test:
Menu Item
Description
Event
Press to display a list of predefined events.
Stage Hold
Press to hold current stage.
Measurements
Reestablish the median complex, set the J point, and select the
three leads used to calculate heart rate.
Leads
Select the leads used for 3 or 6 Rhythm leads, All Leads, Lead
Check, or Lead Placement.
Median
Select a lead to act as the median lead. This can be a fixed lead
or scanned for lead with most ST depression.
Writer
Change the writer’s arrhythmia documentation (doc.), cubic
spline (baseline control), paper speed, gain, filter, and writer
on/off settings.
72
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......