System Setup
Writer Setup
To configure the writer settings for exercise tests, log on to the
System Setup
screen,
select
Exercise Test > Writer Setup
, and complete the fields described in the following
table.
Writer Setup
Field
Description
Speed
Determine the writer's default speed in millimeters per second.
Gain
Determines the writer's default gain setting in millimeters per
millivolts. For example, the setting
10/5
displays limb leads at 10
mm/mV and precordial leads at 5 mm/mV.
Filter
Determines the writer's default filter setting. The default is 150 Hz.
Arrhythmia Doc.
Determines whether a report will print whenever an arrhythmia
occurs during the exercise test.
Tic marks
Enables tic marks on the E, J, and J+ measurement points. The value
for the J+ point is set using the
Post J
field.
ST Measurements
Enables ST Measurements on the display and printout.
Post J
Defines the location, in milliseconds, after the J point where the ST
Measurement is to be taken. Enter a value from 0 to 200.
Writer
Enables or disables the writer for the exercise test.
Exercise Reports
You can configure the reports generated for 12 Lead Exercise/Stress ECGs, 15 Lead
Exercise/Stress ECGs, and Final Report. For each ECG, you can configure the:
• leads used for the reports,
• formats used for the reports, and
• settings for the final report.
In addition, you can configure the extra leads used by the 15 Lead Exercise report.
Report Leads
The
Report Leads
settings allow you to configure the following values:
• the leads used for each standard channel
• the rhythm reporting method
• the leads used for each rhythm report lead group
• the autorhythm lead group
• the extra leads used by the CGR and RMR reports
• the leads used by the Swedish rhythm report
• the method used for the median report
The settings are essentially the same for each ECG type, although the available values
for each setting may differ depending on the ECG type.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
141
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......