System Setup
ECG Reports
You can configure the reports generated for Resting ECGs, Pediatric ECGs, 15-Lead
ECGs, and Vector Loops ECG. For each ECG, you can configure the:
• leads used for the reports,
• report formats used for confirmed ECGs, and
• report formats used for unconfirmed ECGs.
In addition, you can configure the extra leads used by the 15-Lead ECG reports and
the settings for the Vector Loops reports.
Report Leads
The
Report Leads
settings allow you to configure the following values:
• the leads used for each standard channel
• the rhythm reporting method
• the leads used for each rhythm report lead group
• the autorhythm lead group
• the extra leads used by the CGR and RMR reports
• the leads used by the Swedish rhythm report
The settings are essentially the same for each ECG type, although the available values
for each setting may differ depending on the ECG type.
Use the following instructions to select the correct Report Leads for each report type:
•
Resting ECG Report
System Setup > ECG> Resting ECG Reports > Report Leads
•
Pediatric ECG Report
System Setup > ECG > Pediatric ECG Reports > Report Leads
•
15 Lead ECG Report
System Setup > ECG> 15 Lead Reports > Report Leads
•
Vector Loops ECG Report
System Setup > ECG> Vector Loops ECG Reports > Report Leads
Report Leads
Field
Description
Standard leads
Defines which leads will print for each channel. When you change
a channel's lead, the new lead prints on all ECG reports including
the channel.
You can define the lead for channels 1 through 15, except for the
Resting ECG Report, which includes only channels 1 through 12.
Rhythm reports
Defines the data that will print on the rhythm reports:
•
Real time
Prints the ECG data currently on the screen.
•
10 sec delayed
Delays the printing by 10 seconds.
128
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Summary of Contents for MAC 5500
Page 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Page 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Page 187: ......