IMPORTANT: Battery damage may result if the concentrator’s battery is allowed to discharge com-
pletely.
IMPORTANT: After 300 charge/discharge cycles, the battery capacity will be at least 80% of its
original capacity. Replace the battery when the reduced battery life is affecting your mobility.
IMPORTANT: The DC power supply should be utilized on vehicles with the proper power output
rating.
IMPORTANT: The vehicle should be running while utilizing the DC adapter to power the Zen-O lite™
Portable Oxygen Concentrator.
CAUTION:
Remove battery if this device is not going to be used for more than seven days.
Store battery in a cool, dry place.
CAUTION:
Check that this device operates on battery after disconnecting from the power
source.
IMPORTANT: When not using the battery inside the unit, be sure to store it in the protective sleeve
that was provided with the original package. If the battery will not be utilised for long durations,
charge the battery fully at a minimum of once per year.
6.3. Nasal Cannula
Only use a nasal cannula with the following specifications:
•
7ft (2.1 m) or 25ft (7.6 m) long
•
High flow
•
Crush resistant
•
Large internal diameter bore
•
Straight non-tapered tips
•
Suitable for up to 15 litres per minute (lpm) at a max. pressure of 3.6 psi
•
Meets substance compatibility of IEC/EN 60601-1
CAUTION:
Only use approved accessories with this device. Refer to the approved accessories
guide for a complete list of accessories and cannula approved for use with this device. Us-
ing unapproved accessories or cannula may impair the performance of this device, including
flow rate or oxygen purity.
Contact your distributor for updated information and accessories or if additional, optional, or
replacement accessories are needed.