9.1.2. High Priority Alarms
IMPORTANT: These alarms will allow the device to continue operating.
Alarm message
Description
Action
Device is unable to maintain oxy-
gen purity.
Be sure air inlet/outlet has not been
blocked. If alarm persists, contact
your distributor.
Estimated battery life less than 17
minutes.
Charge the battery pack by plug-
ging in to power supply.
Important: The message will be auto-
matically cleared when plugged in to
power supply.
Service required.
Contact your distributor.
*Value: 21-50
9.1.3. Medium Priority Alarms
Alarm message
Description
Action
No breath detected for 15 seconds Check the cannula connection.
Be sure to breathe through nose,
If alarm persists, contact your
distributor.
IMPORTANT: The message will be
automatically cleared when breath-
ing is detected.
Service required.
Contact your distributor.
*Value: 51-70
9.1.4. Low Priority Alarms
Alarm message
Description
Action
Service required.
Contact your distributor.
*Value: 71-99