15.1. Electromagnetic Compatibility (EMC) Information
Medical electrical equipment requires special cautions regarding electromagnetic compati-
bility (EMC). Portable and mobile radio frequency (RF) communications equipment can affect
devices such as the Zen-O lite™ Portable Oxygen Concentrator. As such, the device should not
be used adjacent to other equipment. If this is not practical, then observe the device to make
sure it is operating properly at all times.
15.1.1. Guidance and manufacturer’s declaration: electromagnetic emissions
The Zen-O lite™ Portable Oxygen Concentrator is intended for use in the electromagnetic en-
vironment specified below. The customer or the user of the concentrator should assure that it
is used in such an environment.
Emission test
Compliance
Electromagnetic environment/guidance
RF emissions
CISPR 11
Group 1
The Zen-O lite™ Portable Oxygen Concentrator uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The concentrator is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for do-
mestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies