EN
30/292
Type of protection
(electrical):
Class II
Degree of protection
(electrical):
Type BF
Degree of protection
(water):
IP22 in carry bag (protection against small objects and tilted drip-
ping water)
IP20 out of carry bag (protection against small objects and no
protection against water entering the concentrator)
Degree of safety
(flammable anaesthetic
mixture):
Not suitable for use in the presence of a flammable anaesthetic
mixture
Operating
temperature:
Continuous operation at temperatures between 5°C (41°F) and
40°C (104°F).
Storage temperature:
Between -20°C (-4°F) and 60°C (140°F).
Alarm sound pressure
range:
65 to 85 dB(A)
Alarm system delays:
Less than 10 seconds after detection (low oxygen alarms if oxy-
gen is less than 82% volume fraction at specified environmental
conditions)
Oxygen concentrator
status indicator:
High priority alarm that indicates when oxygen concentration
drops below 82%
Pulse mode bolus size (ml/breath) versus setting and breath rate
Setting
Breath per minute
1
2
3
4
5
15
11
22
33
44
55
20
11
22
33
44
52.5
25
11
22
33
42
42
30
11
22
33
35
35
35
11
22
30
30
30
40
11
22
26.3
26.3
26.3
All /- 15% over all operating conditions
Technical Description (continued)
www.oxygenconcentratorstore.com
877-774-9271