8
SONOSTAT
135
gbo Medizintechnik AG 2004
Version 1.3
2 Introduction
2.1 Transport and installation
The ultrasound therapy device is a portable unit. In the base plate there is a carrying handle in the form of a
recessed grip. Each plane surface is appropriate to place the unit. A wall distance of at least 20 cm has to be
kept. The device must not be placed in front of radiators.
The
SONOSTAT 135
corresponds to the regulations DIN EN 60601. It is a device of protection class I.
Within the scope of the MDD the
SONOSTAT 135
belongs to class IIa .
Warning!
2.2 Connection and switch-on
The ultrasound therapy device is for the connection with mains voltages from 100 - 240 V + 10 %.
Irrespective of the mains voltage the device is appropriate for frequencies of 48 to 63 Hz.
2.2.1 Fuses
1. Unplug the mains plug.
2. The device is protected by 2 fuses that are located in a pluggable box at the back of the device.
3. With a screw driver the box can be pulled out of the receiver by the small slot.
The
SONOSTAT 135
is switched on by the mains switch on the back of the device. By this arrangement
erroneous, unintended disconnection of the device during normal operations shall be avoided.
After switching on the device, a selftest of all functions will be carried out.