SONOSTAT 135
19
gbo Medizintechnik AG 2004
Version 1.3
4.1 Indication menu
An indication menu is integrated in
SONOSTAT 135
. This offers several programs, which suggest certain
parameter settings with regards to ultrasound power, therapy time, operation mode and frequency. These
parameter recommendations are of orientational character only and are subject to the doctors responsibility.
Arthrosis
Calcaneus spur
Carpal tunnel syndrome Cervikal
syndrome Chondropathia patell.
Cicatrix (scars)
Coczygodynia
Coxarthrosis
Distorsion
Dupuytren’s Contrakture
Epikondylitis
Fibrositis
Fracture
Gonarthrosis
Heberden’ disease
Hematonmas, subacute
Intercostalneuralgia
Ischialgia
Lumbalgia
Meniscopathia
Morbus Bechterew
Myalgia
Myogelosis
Neuralgia
Omarthrosis
Periostitis
Rheumatoid-arthritis
Rhizarthrosis
Tendopathia
Tendovaginitis
Trigeminus neuralgia
Sudeck’s dystrophy
1. Switch on the device by the mains switch at the back of the device. Afterwards it carries out an
automatic selftest of all functions. The faultless selftest ends with an acoustic signal (triad gong).
2. After the unit displays the software version it will be operational.
3. Connect one or two ultrasound probes to the socket(s) on the right side of the device.
4. The ultrasound probes will be calibrated. The LEDs in the ultrasound probes flash.
5. After the calibration 05:00 (minutes) are indicated in the display. The LEDs in the ultrasound probes
aren’t lit.
6. Select the desired ultrasound probe by pressing
one of
the
keys
ultrasound
probe small
or
ultrasound
probe large
.
7. Use contact gel liberally onto the area of the patient to be exposed to the ultrasound. Set up the sound
probe flat under light pressure.
8. Select the indication menu by pressing the
softkey
[Ind.]
.
9. There appears a list of indications.
10. Move the scroll bar/cursor with the
modification keys
on the desired indication.
11. Confirm by pressing
the
softkey
[
continue
]
.
12. The indication selected by the indication menu appears with its parameters on the display and the
therapy starts. Except the therapy time and the ultrasound power you can change no parameter.
13. Apply light pressure and keep the ultrasound probe flat against the skin, moving it in circles or even
strokes on the parts to be treated.
14. At the end of the treatment the triad gong will sound 3 x. The ultrasound power is automatically
switched off.