1.3 Short instructions
1. Switch on the device by the mains switch at the back of the device. Afterwards it carries out an
automatic selftest of all functions. The faultless selftest ends with an acoustic signal (triad gong).
2. After the unit displays the software version it will be operational.
3. Connect one or two ultrasound probes to the socket(s) on the right side of the device.
4. The ultrasound probes will be calibrated. The LEDs in the ultrasound probes flash.
5. After the calibration 05:00 (minutes) are indicated in the display. The LEDs in the ultrasound probes
aren’t lit.
6. Select the desired ultrasound probe by pressing
one of
the keys
ultrasound probe small
or
ultrasound
probe large
.
7. The
s
oftkey
[menu]
is operational. This serves for calling the main menu. There you can select the
presettings.
The
softkey
is operational. This serves for loading your own programs. The indication
menu can be selected
with the
softkey
[Ind.]
.
8. In the display the existing parameters are indicated. If wanted you can modify these before the therapy
start.
!!
Note!
The therapy time can also be set during the therapy.
9. Use contact gel liberally onto the area of the patient to be exposed to the ultrasound. Set up the sound
probe flat under light pressure.
10. Increase the ultrasound power with the
intensity regulator
until (normally sensitive) patient has a just
noticeable warm feeling. The therapy starts. The LED in the ultrasound probe are lit.
11. Apply light pressure and keep the ultrasound probe flat against the skin, moving it in circles or even
strokes on the parts to be treated.
12. At the end of the treatment the triad gong will sound 3 x. The ultrasound power is automatically
switched off.
13. Clean the ultrasound probe from the contact gel.