12. Exit the Menu by pressing the
soft
key
[Back]
.
13. On the display the overview window appears again.
Settings
Setting possibilities
Delivery
Acc. coupling control
Activation of the acoustic coupling control, i.e.
with insufficient coupling sounds an acoustic
signal.
on / off
on
Timer coupling control
Linkage between coupling control and therapy
timer. The timer stops if the ultrasound probe is
not coupled.
Without this linkage (set to off) the therapy
time expires continuously after the first
coupling recognition.
on / off
on
Ultrasound power emission
Physical unity of the intensity announcement as a
ultrasound power in W or W / cm2
W or W/cm
2
W
Treatment time
1 - 15
5
Volume gong
0 - 3
3
Volume beeper
0 - 3
3
Display contrast
0 - 9
5
Language
* This language is not activate on delivery.
German
English
Spanish*
Portuguese*
Italian
French
Russian
German
Reset settings
Put back of all settings on manufacturer setting or
delivery condition, with it all individual programs
are also deleted.
Reset settings?
Back / Cont.
Table 1:
Settings