10.5 Current Loop Interface (optional)
Output Type
Isolated active current output 0-20 mA
Non-loaded Voltage
< 20 V
Maximum Output Signal
< 22 mA
Maximum Output Load
(burden resistance):
500 Ohm
Accuracy
Better than 1 %
Connector Type
D-Sub DA-15 female
Used Pins
Current output 1: Pin 9
Current output 2: Pin 2
Return: Pin 1
INFORMATION!
•
Do not connect any other pins.
Current Loop Settings Values (Min, Max, Default and Example Values)
Result Value-Range
Min 0mA, Max 20 mA (Default: 4,0 mA
– 20 mA)
DAC-Range
Depending on Current Loop generation
Error Value
Min 0 mA
– Max 1024 mA (Example: 22 mA –
981 mA)
Idle Value
Min 0 mA
– Max 65535 mA (Example: 20 mA –
54559 mA)