BreathSpec®
– User Manual
155/242
2
Select the import location
pf the sequence file
named
Sequence.json
3
The Sequence will be
executed.
INFORMATION!
The file must be named
Sequence.json
for the software to recognize
it. Make sure, that the parameters specified in the programs and after-
run settings are within your devices permitted regions.
WARNING!
Careful when changing device critical parameters by applying After
Run Settings. A device can become inoperable with faulty IMS
parameters. After Run settings are also applied if the Sequence is
cancelled by the user. In case of cancellation, the devices default
parameters will be the same as if the sequence had successfully
finished, taking all previous After Run Settings into consideration.