BreathSpec®
– User Manual
169/242
3
In
Projection: User Value
-> Current
set
Min
and
Max
value of the
concentration range to the
range 0 mA
– 20 mA.
4
In
Calibration: Current ->
DAC
set
DAC [0 mA]
and
DAC [20 mA]
calibrate to
0-20 mA by observing the
outputted mA value on the
external display unit. The
values on the picture are
examples. The values for
DAC [0 mA] and DAC [20
mA] are preset by G.A.S.
but
must be adapted by
the user.
INFORMATION!
The values for DAC [0 mA] and DAC [20 mA] are preset. The correct
values for DAC [0 mA] and DAC [20 mA] must be adapted by the user.