BreathSpec®
– User Manual
205/242
9.26 Workflow: Manual modification of the sample description
attribute.
The attribute sample description can be changed manually in the measurement file.
For this purpose a text editor is required, e.g. the free editor
+
https://notepad-plus-plus.org/
INFORMATION!
Only authorized, trained and technically instructed people are
allowed to do these work steps.
Please contact us in any case
before you start working.
Only use the recommended text editor +
].
Wrong input can corrupt the measurement file.
Only work with a copy of the measurement file.
1
In windows explorer
select a copy
of the
measurement file.