Action
Repeat the procedure. Restart the system and repeat the
procedure. Check write protection of the USB-Stick. Ensure
that the USB-Stick is FAT32 formatted. Reformat the USB-
Stick. Replace the USB-Stick with a new one. Repeat the
procedure.
If that still does not help contact the G.A.S.
service hotline.
Error message
Do you really want to relabel loop volume? It will be
stored in all following measurements.
Description
Sample loop setting is change manually.
Action
Confirm or abort the dialog.
Error message
Can't set static ip!
Description
Setup of the static IP-address failed.
Action
Check the network settings. Contact your adminstrator.
Repeat the procedure.
If that still does not help contact the
G.A.S. service hotline.
Error message
Can't request DHCP:
Description
Setup of the dynmaic IP-address failed.
Action
Check the network settings. Check DHCP-server. Contact
your adminstrator. Repeat the procedure.
If that still does not
help contact the G.A.S. service hotline.
Error message
Valve set to Inject. Loop not filled
Description
Valve setting is set to position “Inject” and not to “Fill loop”.
Sample loop could not be filled.
Action
Abort program and set valve to
position „Fill loop“ manually.
Error message
Trigger recieved while running program.
Description
During the program run a new trigger signal will be received.
The program run is aborted.
Summary of Contents for BreathSpec
Page 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Page 93: ...BreathSpec User Manual 93 242...
Page 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Page 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Page 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Page 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...