BreathSpec®
– User Manual
118/242
1
Inspect spectrum for
contamination. Control
the baseline. Start
cleaning when
contaminated.
Contamination is
indicated by peaks or the
disappearing of the RIP.
2
Option
Select duration in hours:
System
›
Modes
›
Cleaning Mode
›
x h
3
Option
To speed up this process,
increase flow rate of
EPC1 and EPC2 to their
maximum values (e.g
500/150 ml/min):
Defaults
›
EPC1
›
EPC2