BreathSpec®
– User Manual
231/242
Possible Cause
Sample-Pump has failed
Action
Set pump capacity to 100%. The pump device must be
audible loud. Turn the pumping power back to the working
flow. In addition, the flow can be measured using a flowmeter
at the sample inlet and outlet. If that still does not help
change sample pump.
Symtom
Drift time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating gas.
Action
Check all gas supply connections of GC-column.
Symtom
Retention time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating gas.
Action
Check all gas supply connections of GC-column.