Chapter 8: Measurement of Small Signals--Floating Operation
8 - 7
Ancillary Apparatus
Do not use the Interface 1000 with ancillary apparatus connected directly to any of the cell leads.
Ammeters and voltmeters, regardless of their specifications, almost always create problems when
connected to the Interface 1000 cell leads.
Floating Operation
The Interface 1000 is capable of operation with cells where one of the electrodes or a cell surface is at
earth ground. Examples of earth-grounded cells include: autoclaves, stress apparatus, pipelines, storage
tanks, and battleships. The Interface 1000 internal ground is allowed to float with respect to earth ground
when it works with these cells, hence the name floating operation.
Instrument performance can be substantially degraded when the Interface 1000 is operated in a floating
mode. The instrument specifications only apply on isolated cells with the Interface 1000 earth-ground
referenced (
not
floating).
Take special precautions with the cell connections when the Interface 1000 must float. Make sure that all
the cell connections are isolated from earth ground. In this case, you must disconnect the chassis ground
terminal of the Interface 1000 from earth ground.
Finally, ancillary apparatus connected to the Interface 1000 must be isolated. External voltmeters,
ammeters, FRAs, etc. must be isolated. This includes devices connected to the
Monitor
connector on the
Interface 1000’s front panel.
Caution:
Cables used with the
User I/O
connector must be carefully constructed. We do
not
recommend connection to the
User I/O
connector using a 15-conductor shielded cable. The shield will
generally be connected to the metal shell of the
User I/O
connector on the Interface 1000 end of the cable.
The shield will often be connected to an earth ground on the other end of the cable. Connecting the
grounds in this way destroys the Interface 1000’s ability to float and to make measurements on earth-
grounded cells. In extreme cases, this connection can cause damage to the Interface 1000.
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...