Chapter 5: Panel Indicators and Connectors--Rear Panel
5 - 5
Customization Label Area
The left side of the Interface 1000 front panel has a transparent area in its plastic overlay. A paper label behind
this area allows a custom appearance. This is especially important when the instrument is in a multiple-
potentiostat system.
By default, the label in this area shows the instrument’s serial number written in white text on a blue
background. You can choose a different label. A sheet of alternate labels is provided with each instrument and
a user-editable Excel file can be found on the Gamry website.
Full instructions describing the process of changing a label are near the end of the Installation chapter (Chapter
3).
Rear Panel
Power In Jack
The
Power In
jack is a 4-pin power DIN connector. The Interface 1000 derives all its power from +48 V DC
input through the
Power In
jack. The input current is less than 1.5 A.
Always use the external power supply supplied with your Interface 1000 or power from an Interface Power
Hub to supply DC power to the instrument.
A glowing red
Overload
LED does not necessarily indicate a system malfunction. The
Overload
LED can illuminate when one or more cell leads are disconnected, without indicating a problem
with the system. The
Overload
LED can often illuminate momentarily during a swept or stepped
experiment. The only
Overload
LED indication that definitely points towards a problem is a continuously
glowing
Overload
LED during an experiment.
Customization Label Area
Power In
Jack
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...