Chapter 4: Cell Connections--Membrane Cell Connections
4 - 3
Table 4-2
Cell Cable Connections for ZRA Mode
Color
Type
Name
Normal Connection
Blue
Banana plug
Working Sense
Connect to metal sample #1
Green
Banana plug
Working Electrode
Connect to metal sample #1
White
Pin jack
Reference
Connect to a reference electrode
Red
Banana plug
Counter Electrode
Connect to metal sample #2
Orange
Banana plug
Counter Sense
Connect to metal sample #2
Black
Pin jack
Floating Ground
Leave open or connect to a Faraday shield
The counter sense and the working sense lead are each connected to different metal samples. In the ZRA
mode the Interface 1000 is normally programmed to maintain zero volts between these leads. It therefore
maintains the two metal samples at the same voltage.
The white pin jack on the cell cable is normally connected to a reference electrode. The potential between this
lead and the working sense lead is reported as the cell potential.
In ZRA mode, if you do not have a reference electrode in your cell, we recommend that you connect the white
reference lead to the working electrode. In theory, the measured potential is exactly zero when this is done. In
practice, A/D noise and offset create a small potential signal with a value very close to zero.
Membrane Cell Connections
The Interface 1000 can be used with membrane cells. In this type of cell, a membrane separates two
electrolyte solutions. Two reference electrodes are used: one in each electrolyte. Each electrolyte also contains
a counter electrode. The Interface 1000 controls the potential across the membrane. Table 4-3 shows the cell
connections used with a membrane type cell.
Table 4-3
Cell Cable Connections for a Membrane Cell
Color
Type
Name
Normal Connection
Blue
Banana Plug
Working Sense
Connect to reference electrode #1
Green
Banana Plug
Working Electrode
Connect to counter electrode #1
White
Pin Jack
Reference
Connect to reference electrode #2
Red
Banana Plug
Counter Electrode
Connect to counter electrode #2
Orange
Banana Plug
Counter Sense
Leave open (only needed in ZRA mode)
Black
Pin Jack
Floating Ground
Leave open or connect to a Faraday shield
Reference electrode #1 and counter electrode #1 must be on one side of the membrane,
and reference electrode #2 and counter electrode #2 must be on the other side.
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...