Chapter 5: Panel Indicators and Connectors--Front Panel
5 - 2
The
Power
button is off.
There is no DC +48 V supply connected to the rear panel
Power In
connector.
The external DC power supply has no input power or is malfunctioning.
One part of the Power PC’s power-up self-test has failed.
An Interface 1000’s
Power
LED blinks when that instrument is selected in the Framework’s Instrument
Manager. This allows easy identification of a specific instrument in a Multichannel system without looking at
the instrument Serial Label on the bottom of the instrument chassis.
Finally, the
Power
LED continuously blinks if the instrument is subject to an over-temperature condition. In this
case, the instrument is not usable until it has been powered down, allowed to cool, and restarted.
Cell Cable Connector
The Cell cable connector is a 25-pin D type connector used to connect the Interface 1000 to an
electrochemical test cell. It is normally used with a Gamry Instruments supplied cell cable.
In addition to the pins used for cell connections, the Interface 1000 Cell Connector also uses four pins to read a
cell cable ID. Gamry software can compensate for the cell cable characteristics for optimal system performance,
especially in EIS (Electrochemical Impedance Spectroscopy).
The cell connections are discussed at great length in Chapter 4. A pin-out of the
Cell Cable
connector can be
found in Appendix B.
Monitor Connector
The
Monitor
Connector is a 9-pin DIN-type connector. It contains two analog output signals; one analog input
signal is used for connection of the Interface 1000 to external devices, a temperature input, and an analog
input for measurement of analog input signals. It is not needed for any standard Gamry Framework
electrochemical tests.
Additional information on this connector, its uses, and its signals is in Appendix C.
The USB LED
The
USB
LED is located just below the
Power
LED. It is a tri-color LED, able to glow green, orange, or red.
Caution:
The
Power
LED indicates power status, power-up tests have passed, to identify
instruments, and to indicate latch-up caused by an over-temperature event. Never rely on the
PWR
LED as
a true power-status indicator. Always unplug the
Power In
connection if you suspect your Interface 1000 is
malfunctioning.
Monitor
connector
Cell Cable
connector
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...