Chapter 8: Measurement of Small Signals--Measurement System Model and Physical Limitations
8 - 2
Figure 8-1
Equivalent Measurement Circuit
Icell
C shunt
R shunt
C in
R in
Rm
Unfortunately, technology limits high-impedance measurements because:
Current measurement circuits always have non-zero input capacitance, i.e.,
C
in
> 0.
Infinite
R
in
cannot be achieved with real circuits and materials.
Amplifiers used in the meter have input currents, i.e.,
I
in
> 0.
The cell and the potentiostat create both a non-zero
C
shunt
and a finite
R
shunt
.
Additionally, basic physics limits high-impedance measurements via Johnson noise, which is the inherent
noise in a resistance.
Johnson Noise in
Z
cell
Johnson noise across a resistor represents a fundamental physical limitation. Resistors, regardless of
composition, demonstrate a minimum noise for both current and voltage, per the following equations:
E
= (4
kTR
F
)
1/2
I
= (4
kT
F
/
R
)
1/2
where
k
= Boltzman’s constant, 1.38 × 10
−23
J/K
T
= temperature in K
F
= noise bandwidth in Hz
R
= resistance in Ω.
For purposes of approximation, the noise bandwidth,
F
, is equal to the measurement frequency. Assume a
10
11
Ω resistor as
Z
cell
. At 300 K and a measurement frequency of 1 Hz, this gives a voltage noise of 41 µV
rms. The peak-to-peak noise is about five times the rms noise. Under these conditions, you can make a
voltage measurement of
10 mV across
Z
cell
with an error of about
0.4%. Fortunately, an AC
measurement can reduce the bandwidth by integrating the measured value at the expense of additional
measurement time. With a noise bandwidth of 1 mHz, the voltage noise falls to about 1.3 µV rms.
Current noise on the same resistor under the same conditions is 0.41 fA. To place this number in
perspective, a
10 mV signal across this same resistor generates a current of
100 fA, or again an error of
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...