Appendix A: Interface 1000 Specifications--Floating Operation
Appendix A - 2
Current to Voltage Converter
Maximum Full-scale Range
1000
mA
Note 13
Minimum Full-scale Range
10
100 (after ×100 gain)
nA
pA
Note 13
Voltage across Rm
Typ
100
mV at
full scale
Note 14
Output Voltage (at Monitor and ADC in)
Typ
3.0
V
full scale
Input Offset Current
Max
10
pA
Note 15
Range Zero Offset
Max
0.05
% of
range
Note 15
Gain Tolerance
1 A to 10 nA ranges
Max
0.3
% of
reading
Note 15
Zero drift
Typ
0.03
% FS /
C Note 16
Bandwidth
Typ
>1.5 MHz (10 µA)
>0.15 MHz (1 µA)
MHz
Note 17
Current Measurement
Resolution
Typ
0.00333…
% FS
/bit
Offset Range
Typ
±100
% of
range
Post-Offset Gain
Typ
1×, 10×, 100×
Accuracy
Typ
Dominated by current-to-voltage error
(see above)
Summary of Contents for Interface 1000
Page 8: ......
Page 18: ......
Page 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Page 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Page 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Page 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Page 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Page 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Page 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Page 80: ...Appendix E CE Certificate Appendix E 4...
Page 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...