15.
Technical specifications
Dimensions (maximum)
472 X 432 X 131 mm, length x width x height (fits standard
19-inch wide device rack)
Shipping weight (maximum) 32 (66) kg (lbs) (fully configured with two supplies)
Operating temperature
0 to 40 °C Free airflow around the unit is required.
Altitude
Sea level to (2000 m) 6560 ft.
Humidity
0 to 80% RH (non-condensing)
Storage temperature
10 to 70 °C
Input voltage
110 V a.c. or 220 V a.c., +/- 10%. (input mains voltage
configuration specified at order time)
Input current
10 A fused
Line frequency
50/60 Hz
High Voltage Supply
▪ Power supply type
Linear (HV transformer)
▪ Number of channels
(maximum)
2
▪ Voltage (fixed output)
Default configu 7 kV (CV/DI) or - 5.6 kV (TR)
▪ Voltage resolution
100 V
▪ Current per channel
(maximum)
500 mA (short circuit)
▪ Current resolution
(maximum)
0.1 uA
▪ Power per channel
(maximum)
500 W
▪ Output connector type
SHV-10 or Fischer Type (other connector types are
available per request)
▪ SAFECONN Connector
(HV Interlock) SMB (with SHV-10 output connector) or Mini
Banana (with Fisher output connector)
Display or controls
7” TFT WVGA (800 x 480) Colour LCD, 16:9 aspect ratio,
Touchscreen
Communications
Ethernet, Serial (RS-232, RS-485HD 2 wire mode, RS-485FD
4 wire mode), Profibus (optional)
Available analog and digital I/O connections
Setpoint relays
4 relays, 500 mA maximum, 28 V maximum, user
configurable
Setpoint logic outputs
4 TTL Outputs, 6 mA maximum, 5 V, user configurable
Logic inputs
4 TTL inputs, internally pulled up to 3.3 V, user
configurable
Analog outputs
4 analog outputs 0 - 10 V, user configurable (voltage,
current or pressure)
Page 51
900034_C
900034_C - Technical specifications