Hex
command
Description
Data
field
Response Data and response interpretation
5A
HV GET/SET ANALOG
OUTPUT
Configures analog
outputs.
If only N parameter is
specified, the
current analog
output configuration
is
returned.
N, S, F, O,
I, R
S, F, O, I,
R
Where:
N is analog channel (1-4)
S is pump (1-4)
F is function (see below)
O is log offset (-15 to +15)
I is inverted mode (0-normal, 1-
inverted)
R is response mode (0-normal, 1-
fast)
Function:
“00” - Off
“01” - Logarithmic pressure
“02” - Logarithmic current
“03” - Volts per 1 uA
“04” - Volts per 10 uA
“05” - Volts per 100 uA
“06” - Volts per 1 mA
“07” - Volts per 10 mA
“08” - Volts per 50 mA
“09” - Volts per 1 nA
“10” - Volts per 10 nA
“11” - Volts per 100 nA
“12” - Volts per 1 kV
28
TSP TURN OFF
Turns off TSP
sublimation.
R
Where:
R is TSP (1-2)
29
TSP SET ACTIVE
FILAMENT
Sets TSP active
filament.
R, N
Where:
R is the TSP (1-2)
N is the filament number (1-4)
2D
TSP TURN ON
Turns on TSP
programmed mode. R
Where:
R is TSP (1-2)
DF
TSP GET ACTIVE
FILAMENT
Reads active TSP
filament.
R
N
Where:
R is TSP (1-2)
N is the active filament.
Page 45
900034_C
900034_C - Serial communication