5.
Controller configuration
Following configuration options are specified at order time:
5.1.
Input mains AC voltage (110 V a.c. vs. 220 V a.c.)
Controller does not have universal power input, thus input mains AC voltage
is part of the controller configuration.
Care must be taken when connecting controller to mains AC voltage to avoid
connecting controller to wrong input voltage.
5.2.
Supply configuration
Following controller configurations are valid:
1.
One HV supply or channel
2.
Two HV supplies or channels
3.
One HV supply or channel and one internal TSP supply
5.3.
Output voltage polarity for each HV channel (positive vs negative)
In case of positive polarity, output voltage is set to 7 kV by default.
In case of negative polarity, output voltage is set to -5.6 kV by default.
Output voltage is fixed (linear power supply, transformer tap).
5.4.
Type of the HV output connectors
Following are standard HV connector types for J501 - J504:
1.
SHV-10 kV (matching safeconn connector type is SMB style).
2.
Fischer (matching safeconn connector type is mini banana style) other HV
connector types are available per request.
5.5.
Number of HV output connectors per HV supply
There can be more than one HV output connector tied to the same HV
supply. In such case, connectors are tied in parallel.
Controller can have maximum of 4 HV output connectors in any
configuration. For example, 2 HV output connectors per 1 HV supply.
5.6.
Communication interface
Controller comes standard with Serial and Ethernet interface.
Profibus interface is optional.
5.7.
Remote TSP control
Controller comes standard with connector J505 CTRL installed.
Connector can be used to connect and control remote TSP Controller.
See rear panel section for connector location.
If internal TSP supply is installed, connector J505 is not available.
Page 14
900034_C
900034_C - Controller configuration