‘TSP Armed’ - TSP is running programmed mode, but not currently
sublimating TSP filament.
‘TSP Armed, Supply is off’ - TSP cannot continue because pressure window is
set, but supply is off.
‘TSP Armed, Outside the pressure window’ - TSP cannot continue because
pressure window is set, but pump pressure is outside the pressure window.
‘TSP Armed, Interlock - Active’ - TSP cannot continue because TSP interlock is
set, but not satisfied.
‘TSP Armed, Waiting – In use’ - TSP cannot continue because other TSP is
currently running the sublimation cycle.
Figure 17. TSP sublimation view screen (no control buttons)
9.15.
TSP sublimation view with controls
View pump - Switches home screen to pump view.
Menu - Invokes main menu.
Exit - Clears screen and hides control buttons.
Turn On TSP (Turn Off TSP) - Starts TSP programmed mode according to the
parameters defined in TSP setup window. If TSP programmed mode is
running this button will turn off TSP operation.
TSP setup - Invokes TSP setup screen.
Page 28
900034_C
900034_C - Touchscreen user interface