![Gamma Vacuum DIGITEL MPCq Instruction Manual Download Page 34](http://html1.mh-extra.com/html/gamma-vacuum/digitel-mpcq/digitel-mpcq_instruction-manual_3377554034.webp)
Note:
Connector is not installed if controller is configured with internal TSP supply.
11.4.
Connector J114, ETH
Standard RJ-45 connector. Connection to Ethernet interface.
Connector has two LED indicators
▪
Green LED - Indication of link state.
▪
Orange LED - Indication of data activity.
11.5.
Connector J510, TSP FIL (TSP, optional)
Installed in case when MPCq is configured with internal TSP supply.
Three filament, keyed TSP Pump connector.
Use high current cable to connect to Titanium Sublimation Pump (TSP).
11.6.
Connector J104, Misc I/O
Standard Sub-D 37-pin female connector.
Table 2. J104 connector pinout
Pin
Description
1
Relay 1 Common
*
2
Relay 1 NC
3
Relay 1 NO
4
Relay 2 Common
*
5
Relay 2 NC
6
Relay 2 NO
7
Relay 3 Common
*
8
Relay 3 NC
9
Relay 3 NO
10
Relay 4 Common
*
11
Relay 4 NC
12
Relay 4 NO
13
Reserved - do not use, ground
14
Reserved - do not use, ground
15
DIGITAL OUT 1 - output, digital, 0/+ 5 V d.c., 7 mA maximum
16
Reserved - do not use, ground
17
DIGITAL OUT 2 - output, digital, 0/+ 5 V d.c., 7 mA maximum
18
Reserved - do not use, ground
19
DIGITAL OUT 3 - output, digital, 0/+ 5 V d.c., 7 mA maximum
20
Reserved – do not use, output, digital, 0/+ 5 V d.c., 7 mA
maximum
Page 34
900034_C
900034_C - Rear panel