Figure 18. TSP sublimation view screen with controls
9.16.
TSP setup
Available setup:
Filament - Selects active filament (1 - 4). Filament status reads (OK) if at the
end of the firing cycle, filament current reached 80% of current target value.
A question mark (?) is
displayed if filament current did not reach 80% of the current target value.
Current target (A) - Configure sublimation current target level.
On time (s) - Configures sublimation time in seconds.
Number of cycles - Configures number of sublimation cycles to run.
Cycle period - Configures time in minutes between each sublimation cycle.
Advance options configure - Invokes TSP advance options window.
TSP 1 and TSP 2 - Selects TSP for which above parameters are configured.
Figure 19. TSP setup screen
Page 29
900034_C
900034_C - Touchscreen user interface