Hex
command
Description
Data
field
Response Data and response interpretation
34
HV GET SUPPLY AUTO
RESTART
Reads supply auto
restart option.
S
YES or
NO
Where:
S is supply (1-2)
37
HV TURN ON
Turns on supply HV.
S
Where:
S is supply (1-2)
38
HV TURN OFF
Turns off supply HV.
S
Where:
S is supply (1-2)
3B
HV GET/SET SET-
POINT
Configures setpoints.
Set-points (N
parameter) 1-4 are
relays. Set-points (N
parameter) 5-8 are
digital outputs.
If only N parameter is
specified, the current
set-point
configuration and
state is returned.
N, F, S,
X.XE-XX,
Y.YE-YY
N, F, S,
X.XE-XX,
Y.YE-YY, A
Where:
N is the set-point number (1-8)
F is the function 0=Off, 1=Pressure
Setpoint,
2=HV Error, 3=HV On/Off Indicator
S is the configured supply (1, ,2 ,3
or 4)
X.XE-XX is the On Pressure
Y.YE-YY is the Off Pressure*
A indicates: 1 = Relay/TTL output
energized,
0 = Relay/TTL output de-
energized)
*Off Pressure must be at least
20% greater than On Pressure. If
not, controller will automatically
set Off Pressure to 20% greater
than On Pressure.
58
HV GET/SET DIGITAL
INPUT
Configures digital
inputs. If only N
parameter is
specified, the current
digital input
configuration is
returned.
N, S, F
S, F
Where:
N is the digital input (1-4)
S is the pump (1-2)
F is function:
“00” - Off
“01” - HV Interlock
“02” – HV Switch On/Off
“03” – TSP 1 Interlock
“04” - TSP 2 Interlock
Page 44
900034_C
900034_C - Serial communication