Digitel MPC Manual
Page 50 of 72
2F
DEGAS. A maintenance tool. that conditions your filaments. Each filament is fired once. DEGAS uses
the specified ON TIME and SUBLEVEL setup parameters, but starts its firing at half the value. For
example, if current is specified at 50 amps and time at 30 seconds, then the TSP fires at 25A for the first
15 sec and then steps up to 50A for the remaining 15 sec.
32
CONTROL FAN. Parameters: ON or OFF.
33
SET AUTO-RESTART. Sets pump to automatically restart on power up. Parameters: YES or NO.
34
GET AUTO RESTART. Response: YES or NO.
37
START PUMP. Starts specified pump. Parameters: 1 or 2*.
38
STOP PUMP. Stops specified pump. Parameters: 1 or 2*.
3C
GET SETPOINT. Reads configuration of specified set point. Response:
n
,
s
,
X.XE-XX
,
Y.YE-YY, ST
.
n
= set point number (1-8);
s
= supply driving it;
X.XE-XX
= on pressure;
Y.YE-YY
= off pressure, ST
indicates current setpoint state, 1 = ON, 0 = OFF.
3D
SET SETPOINT. Configures a specified set point. Parameters: n, s, X.XE-XX, Y.YE-YY.
n = set point number (1-8); s = supply driving it (0=inactive, 1=supply 1, 2=supply 2, 3=bake supply 1,
and 4=bake supply 2); X.XE-XX = on pressure; Y.YE-YY = off pressure.
3E
SETUP BAKE. Arms the bake. Parameters: hhh, HH:MM. hhh is bake duration (0 –255 hours); HH:MM
is the start time in hours and minutes.
3F
BAKE TIME LOG. Response:
nnn.nH
.
nnn.nH
is the total time the set points have been on.
40
DISABLE BAKE.
44
LOCK KEYPAD.
45
UNLOCK KEYPAD.
50
GET ANALOG OUTPUT MODE. Gets the current/pressure analog output mode for each supply.
Parameters 1 or 2*. Response: “0 Logarithmic pressure”, or "1 Logarithmic current", or "2 Volts per
1.0uA", or "3 Volts per 10.0uA", or "4 Volts per 100.0uA", or "5 Volts per 1.0mA", or "6 Volts per 10.0mA",
or "7 Volts per 50.0mA".
51
SET ANALOG OUTPUT MODE. Sets the current/pressure analog output mode for each supply.
Parameters 1 or 2*, n. Where n = 0 – 7. 0 = Logarithmic pressure, 1 = Logarithmic current, 2 = 1uA, 3 =
10uA, 4 = 100uA, 5 = 1mA, 6 = 10mA, 7 = 50mA.
C6
READ EVENT LOG. Reads event data from the event log.
Parameter: 0 – 2048. Events are numbered 1 – 2048. 0 will return just the latest event.
Response: n string
where n is the event number
string is the following format
pump# error code
time & date
Voltage current
pump size supply size
# of arcs # of cool down cycles
0 0 0 0 (data reserved for future)
* Parameters 1 or 2 allow you to select which supply.