1.
Press
ENTER
. An arrow points to the top
Display Select
key to show it is active.
Press it to toggle between AMPS and WATTS.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
arrow shows
bottom key is active
Options:
- AMPS
- WATTS
press to toggle
between options
2.
When you reach the desired option, press
ENTER
.
3.
The bottom display shows the value followed by a flashing sign. Use the backspace
(
←
) key to overwrite, and enter the desired value (0-75 for amps or 0-500 for watts)
with the numerical keys.
1
2
3
4
5
6
7
8
Set Point
Set Point
50
Supply 1
Supply 2
Display
Select
Display
Select
flashing
sign
Options:
- 0 to 75 amps
- 0 to 500 watts
4.
Press
ENTER
to save your changes.
AUTO--ADVNC
AUTO--ADVNC, when set to YES, automatically switches to the next filament if an open
filament is detected.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
When in TSP SETUP, press the Menu key to cycle to AUTO--ADVNC.
1.
Press
ENTER
. An arrow points to the bottom
Display Select
key to show it is
active.
AUTO--ADVANCE
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
arrow shows
bottom key is active
Options:
- YES
- NO
press to toggle
between options
2.
Press it to toggle between YES and NO.
3.
When you reach the desired option, press
ENTER
.
Menu Key
The
Menu
key (blank key) is a hierarchical system (Table III-3) for setting unit parameters.
After you press
Menu
, the top display shows the first menu selection.
Digitel MPC Manual