2.
An arrow points to the bottom
Display Select
key to show it is active. Press it to
cycle through the options: F1 ACTIVE, F2 ACTIVE, F3 ACTIVE, F4 ACTIVE.
3.
When you reach the desired filament, press
ENTER
to save your selection or press
ESC
to exit without saving changes.
CLR FILMNT
If you degas filaments or install new ones, use CLR FILMNT to reset their status to OK (not
open) and their operating logged time to zero. Entering 0 resets all four filaments.
When in TSP SETUP, press
Menu
to cycle to CLR FILMNT.
Press
ENTER
. When the display asks you to confirm, press
ENTER
to clear the filaments or
press
ESC
to exit without clearing filaments.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
ON TIME
This parameter specifies the on firing time of a filament for TIMED and DEGAS operating
modes. For example, if you specify 30 sec ON TIME (in TIMED mode), then the TSP operates
for 30 sec before it shuts off.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
When in TSP SETUP, press
Menu
to cycle to ON TIME.
1.
Press
ENTER
. The bottom display shows the on time in seconds followed by a
flashing sign.
TSP SETUP
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
flashing
sign
Options:
0 through 999 sec
2.
Use the backspace (
←
) key to overwrite, and enter the desired value (0-999) with the
numerical keys. Press
ENTER
to save the changes.
SUB LEVEL
For TSP control, you need to select constant current (amps) or constant power (watts) and
specify its value.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
When in TSP SETUP, press the
Menu
key to cycle to SUB LEVEL.
Digitel MPC Manual