1.
3-wire, detachable AC input power cable (included with DIGITEL)
2.
high voltage (HV) cable for each pump (ordered separately)
3.
safety ground cable for each pump (ordered separately)
Procedure
Perform the following procedure to install the MPC:
1.
Place the unit in its location and secure as necessary.
2.
Connect the safety ground cable between the pump and safety ground stud at the
DIGITEL rear panel (Figure II-1).
3.
Connect the high voltage cable to the ion pump and to the appropriate high voltage
jack on the DIGITEL rear panel. See Figure II-1 and Table II-1.
4.
If you have the optional SAFE-CONN feature, an additional connector is part of the
HV connector. Connect it to the appropriate SAFE CONN jack. See Figure II-1 and
Table II-1.
5.
Verify correct input voltage requirements. Connect the AC input power cable to
input power receptacle on the DIGITEL rear panel.
6.
Switch on the DIGITEL and set correct pump size.
HV connect - supply 1
safety ground cable
SafeConn connect - supply 1
Optional
SafeConn connect - supply 1
Optional
HV connect - supply 1
ac
input
Figure II-1. MPC Rear Panel HV and SAFECONN Connectors.
Table II-1. High Voltage and SAFE-CONN Connectors.
HV Connector
SAFE-CONN Connector
Description
J501
J401
HV & SafeConn Connect, Supply 1
J502
J402
Optional Connect, Supply 1
J503
J403
HV & SafeConn Connect, Supply 2
J504
J404
Optional Connect, Supply 2
Installing the Remote TSP/NEG Control Option