SPLY1 SIZE / SPLY2 SIZE
To display the supply option installed (large, medium or none) for supply 1 and supply 2, use the
following procedure.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to SPLY1
SIZE.
3.
Press the
ENTER
key. The bottom display shows the pump option installed for
supply 1.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
Options:
- LARGE
- MEDIUM
- NONE
4.
Press the
Menu
key twice to select SPLY2 SIZE, and press
ENTER
. The bottom
display shows the pump option installed for supply 2.
PUMP 1 CAL FACTOR / PUMP 2 CAL FACTOR
Each pump has a calibration factor. This factor is multiplied by pressure to increase or decrease
it. You need to specify an acceptable factor between 0.00 and 9.99. Use the following
procedure.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to PUMP 1
CAL FACTOR.
3.
Press the
ENTER
key. The bottom display shows the calibration factor with the first
number flashing.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
Number flashes.
Enter cal using
the keypad.
4.
Key in a new value and press
ENTER
to save the changes.
5.
Press the
Menu
key to select PUMP 2 CAL FACTOR.
6.
Repeat Steps 3 and 4.
SUPPLY1 KV / SUPPLY2 KV
To display what voltage is strapped to the high voltage transformer for supply 1 and 2 (5600 or
7000), use the following procedure.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to SUPPLY1
KV.
3.
Press the
ENTER
key. The bottom display shows the voltage strapped to supply 1.