Digitel MPC Manual
Page 42 of 72
Front Panel Display of Events
Each event is displayed in five parts (Table III-6). Each part files the dual 10 character display.
Table III-6. The Five Parts of the Event Record.
Part Format
Description
1.
EVNT n >
e is the event number
PMP n ee <
n is the effected pump, and ee is the error code (see section ERROR CODE on page 39)
2.
mm/dd/yy >
mm is month, dd is day, yy is year that the event occurred
hh mm ss <
hh is hour, mm is minute, ss is second that the event occurred
3.
nnnn V > nnnn is the high voltage when the event occurred
n.nn –ee A < current when the event occurred
4.
xxx nn >
xxx is either LRG or MED for supply size, nn is pump size in liters per second
n c <
n is number of arcs, c is number of cool down cycles
5.
n 0 >
is the pump state, 0 is data reserved for future use
0 0 <
for future use
Serial Port Output of Events
The serial port merely combines both display lines into a single line, and puts out 5 lines.