Installing the Cell Processing Set
COBE 2991 Cell Processor
•
Essentials Guide
6-13
P
rocessing
and C
o
mmon
Operations
6
28
Verify that the rubber collar of the rotating seal is pushed all the way into the
circular slot at the rear of the seal weight (Figure 6-12).
Figure 6-12: Correct and incorrect rotating seal installation
29
Close the centrifuge cover interlock latch (if applicable) by rotating the latch
knob fully counterclockwise.
30
Do one of the following:
•
If the READY light is illuminated, the processing set is now installed and
ready for use. Proceed to “Loading the Starting Cell Product” on page 6-14.
-or-
•
If the READY light is not illuminated, press the STOP RESET button to exit
the Tube Load mode. If the READY light still does not illuminate, ensure the
sliding covers are closed, the centrifuge cover interlock latch is closed (if
applicable), and the hydraulic pump is at its home (starting) position. If the
READY light remains off, proceed to “Ready Light Does Not Illuminate” on
page 7-13.
Caution:
Do not interfere
with the seal
Correct
Incorrect
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...