4-6
COBE 2991 Cell Processor
•
Essentials Guide
LED Programming
Programming a Hold at the Beginning of the Agitate/Wash-In Operation
To program a Hold status at the beginning of the Agitate/Wash-In operation for a
given cycle, do not program a valve for that cycle.
When the Agitate/Wash-In operation with no valve programmed is reached, the
HOLD light illuminates, the centrifuge stops, the primary audible alarm sounds (if
turned on), the secondary audible alarm sounds, and the MINIMUM AGITATE
TIME timer is temporarily suspended.
You have the following options:
•
Program the desired valve to open. Press the CONTINUE button. The
centrifuge agitates, the selected valve opens, and the Agitate/Wash-In
operation continues as programmed.
-or-
•
Do not program a valve to open. Press the CONTINUE button. The
centrifuge agitates, and the Agitate/Wash-In operation continues as
programmed without processing/wash solution being added.
For a summary of CALL stop (Hold) and STOP SELECT options available with
the LED program board, see the following table.
Table 4-1: Call/Stop Matrix
Option
Selected
When
Call/Stop
Occurs
Audible
Alarm?
Centrifuge
Status
Valve
Status
Next Action
CALL stop
Beginning of
Super Out
Yes (both)
Spinning
SOV or
Collect open
Press
CONTINUE
.
P.C. stop
After Super
Out
Yes
(primary
only)
Stopped
Closed
Press
STOP RESET
.
R.C. stop
After
Agitate/Wash-
In
Yes
(primary
only)
Stopped
Closed
Press
STOP RESET
.
No valves
programmed
in cycle.
Beginning of
Agitate/Wash-
In
Yes (both)
Stopped
Closed
Program
valve and/or
press
CONTINUE
.
Zero spin
time
programmed
in cycle.
Beginning of
Start/Spin
Yes (both)
Spinning
Closed
Enter time
≥
0.1 minute
and press
CONTINUE
, or
press
STOP RESET
.
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...