1-24
COBE 2991 Cell Processor
•
Essentials Guide
Introduction
Agitate/Wash-In Operation
The Agitate/Wash-In operation (Figure 1-18) follows the Supernatant-Out
operation. The centrifuge decelerates to a near stop. The hydraulic pump reverses to
its home (starting) position, removing hydraulic fluid from underneath the flexible
diaphragm. The selected valve opens and the centrifuge begins to agitate.
Processing/wash solution flows by gravity into the cell processing bag as the
oscillating motion of the centrifuge causes the cell product and the processing/wash
solution to mix. The COBE 2991 advances from the Agitate/Wash-In operation to
the next cycle when
•
The hydraulic pump has returned to its home (starting) position.
-and-
•
The selected MINIMUM AGITATE TIME has elapsed.
-or-
•
The START SPIN button is pressed.
This concludes the first cycle.
See Table 1-6 on page 1-25 for a summary of the Agitate/Wash-In operation.
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...