Table 5-4: Function of Each Tab
Screen Name
Function
Analysis
Allows the user to see the current status of the analyzer, including
concentration readings and input and output status.
Status
Provides the current status of the analog outputs, relays, solenoids,
analog inputs, discrete inputs, and what voltage input source is in use.
Alarms
Allows the operator to view all enabled system alarms.
Archive
Allows the operator to access the numerous archives that store live data,
hourly and daily averages, as well as other user configurable archives.
Events
Allows the operator to access the
Event Log
, which keeps track of all the
configuration changes and events that the analyzer has experienced.
Global
Allows the operator to change settings that affect the global operation of
the analyzer, such as stream purge times.
Streams
Allows the operator to change settings specific to each stream; normal,
zero, reference, and calibration.
Inputs-DRY
Allows the operator to configure the four discrete (Dry) inputs of the
analyzer.
Inputs-WET
Allows the operator to configure the four discrete (Wet) inputs of the
analyzer.
Outputs
Allows the operator to configure the behaviour of the relays, solenoids and
analog outputs.
Sequences
Allows the operator to configure the analyzer gas run event sequences in
either, automatic or manual or both.
Ports
Allows the operator to view the settings of the analyzer’s communications
ports - RS-232, RS-485, and USB.
Modbus
Allows the operator to configure Modbus lists, including the type of
Modbus list and the data points that are outputted to Modbus.
Analog inputs
Allows the operator to configure the 4 analog inputs including the live
pressure and temperature readings and related alarm set points.
Timers
Allows the operator to schedule conditional gas run operational requests
on an hourly, daily or weekly basis.
Maintenance
Allows the operator and service personnel to have greater control over the
electronics board, for testing analog outputs, relays, and solenoids.
Factory
Reserved for factory and service personnel
5.2.4 Status Bar
The
Status
bar, which is located at the bottom of the window, presents the update status, the
present time/date and traffic lights to indicate the status of communication between the
analyzer and the computer.
Summary of Contents for ProTech903
Page 2: ......
Page 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Page 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Page 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Page 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Page 190: ...Revision 14 190 August 02 2019...