H2S-LC 4.00 OFF
- This line is used to indicate if the Lock Stream alarm should be enabled
for the H2S measurement to change the setting, press the EDIT button. The operator will be
prompted to enter the desired value and press the ENTER key on the keypad.
•
TS-H 16.00 OFF
- This line indicates the set point for the Total Sulfur High alarm and the
current status of the alarm. The alarm set point can be edited by pressing the EDIT key on
the keypad while this line is highlighted. A new value can be typed in via the numerical keys
and pressing the ENTER key will cause the new value to be accepted.
•
TS-HH 16.00 OFF
- This line indicates the set point for the Total Sulfur High High alarm and
the current status of the alarm. The alarm set point can be edited by pressing the EDIT key
on the keypad while this line is highlighted. A new value can be typed in via the numerical
keys and pressing the ENTER key will cause the new value to be accepted.
•
TS-DUR 16.00
- This line indicates the length of time in seconds that the Total Sulfur High
High alarm set point has to be exceeded before the alarm will be activated. This parameter
can be edited by pressing the EDIT key on the keypad while this line is highlighted. A new
value can be typed in via the numerical keys and pressing the ENTER key will cause the
new value to be accepted.
•
TS-LCK 4.00 OFF
- This line is used to indicate if the Lock Stream alarm should be enabled
for the T measurement. The value is set via the GUI (Section 5.10.3.6). To change the
setting, press the EDIT button. The operator will be prompted to enter the desired value and
press the ENTER key on the keypad.
•
GAIN 100.0 100.0
- This line is used to set the gain. To change the setting, press the EDIT
button. The operator will be prompted to enter the desired value and press the ENTER key
on the keypad. In
Edit
mode, the left value changes while the right value stays at the
previous value (in case you cannot recall what the previous value was).
•
CALGAS 18.0 18.0
- This line is used to indicate the concentration of the calibration gas. To
change the setting, press the EDIT button. The operator will be prompted to enter the
desired value and press the ENTER key on the keypad. In
Edit
mode, the left value changes
while the right value stays at the previous value (in case the operator cannot recall what the
previous value was.
•
OFFSET 0.0 0.0
- This line is used to indicate the current value of the zero offset. To
change the setting, press the EDIT button. The operator will be prompted to enter the
desired value and press the ENTER key on the keypad. For additional information, see
Section 5.10.4.3. In
Edit
mode, the left value changes while the right value stays at the
previous value (in case the operator cannot recall what the previous value was.
•
OGAS 0.1 0.1
- This line is used to indicate the concentration of H
2
S in the zero gas. To
change the setting, press the EDIT button. The operator will be prompted to enter the
desired value and press the ENTER key on the keypad. In
Edit
mode, the left value changes
while the right value stays at the previous value (in case the operator cannot recall what the
previous value was.
•
CAL OFF
- This field is used to initiate a calibration run. To initiate the calibration run
press the EDIT button, and press ENTER to toggle the status from OFF to ON. The analyzer
will run the calibration according to the setup of the Calibration run (see Section 5.10.4) and
then revert back to the auto-sequence.
•
REF1 OFF
- This field is used to initiate the Reference 1 run. To initiate the Reference
1 run press the EDIT button, and press ENTER to toggle the status from OFF to ON. The
analyzer will run the Reference 1 run according to the setup of the Reference 1 run (see
Section 5.10.4) and then revert back to the auto-sequence.
•
REF2 OFF
- This field is used to initiate the Reference 2 run. To initiate the Reference
2 run press the EDIT button, and press ENTER to toggle the status from OFF to ON. The
analyzer will run the Reference 2 run according to the setup of the Reference 2 run (see
Section 5.10.4) and then revert back to the auto-sequence.
Summary of Contents for ProTech903
Page 2: ......
Page 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Page 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Page 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Page 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Page 190: ...Revision 14 190 August 02 2019...