Revision 14
178
August 02, 2019
11.4 Instrument Specifications
Size
28”Hx16”Wx16”D (71 cm x 41 cm x 41 cm)
Weight
Class 1 Div 1 80 lb 36.4 kg
Class 1 Div 2 40 lb 18.2 kg
Power Consumption
10W @ 24VDC
250W @ 120 AC (Total sulfur option)
Power Input Dual power design - Universal 90-
230VAC
and 10-36VDC input. Capable of being
attached
to redundant power supply
Electrical
CSA C/US Certified
Classifications
Class 1, Div 1, Group B,C&D
Class 1, Div 2, Group B,C&D
Ambient Temperature
0-50°C (without enclosure)
Ambient Humidity
0 – 95% non-condensing
Electronics
Dual processor design with intelligent
colorimetric sensor. To achieve greatest
sensitivity sensor uses 24 bit AD converter
equipped with temperature monitor on sensor
block. Sensor automatically calibrates on every
cycle to prevent drift.
Optical tape counter optimizes tape usage and
provides exact measurement of tape used.
User can program low tape alerts/alarms.
Real time clock for accurate event time
stamping.
Software
Windows based software for configuration and
monitoring. Software includes advanced
configurable features for stream switching,
automatic calibration, multi mode H2S and total
sulfur measurement, reference and zero gas
test, and multi-ranging.
Humidifier
Permeable membrane humidifier system
facilitates faster response time and eliminates
moisture carry-over to lead acetate tape.
Memory Field programmable memory
Firmware for unit can be upgraded in the field.
Data Logging Up to 10 months of H2S
concentration data logs. 512 Kbytes non-
volatile memory for data logging and audit trail.
Options
Total sulfur system, dilution system, high
pressure sampling system, solar power, back
panel, low pressure sensor, low hydrogen
sensor, auto-cal, and enclosure / cabinets
(temperature
Ranges
Up to 0-500ppm with no dilution; up to 0-100%
with permeation dilution system
Low Range
0-1ppm or less (H
2
S or total sulfur). Please
consult factory for more information.
Summary of Contents for ProTech903
Page 2: ......
Page 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Page 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Page 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Page 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Page 190: ...Revision 14 190 August 02 2019...