Revision 14
74
August 02, 2019
5.2.2.2 Operation Toolbar
The
Operation
toolbar contains several buttons that control various aspects of the operation
of the analyzer and the GUI. Table 5-2 shows the buttons in the
Operation
toolbar.
Table 5-2: Operation Toolbar Buttons
Button
Name
Function
Temporary Write
to Analyzer
Writes any configuration changes in the active screen
to the analyzer’s volatile memory.
Permanent Write
Configuration to
Analyzer
Writes any configuration changes in the GUI first to the
volatile memory and then writes the changes to non-
volatile flash memory. Any changes that the user
wishes to be permanent and to be present if the
analyzer is powered off must be written to flash. If the
analyzer is powered down before changes are written
to flash, those changes will be lost.
Poll Page Once
Reads back all the data points on the current screen.
Pushing this button allows the user to check and
ensure that any changes written to the analyzer have
been accepted.
Synchronize
Time
Updates the on-board real time clock in the analyzer to
the time and date set on the linked computer.
Acknowledge All
Latches
Returns all latches that may have been triggered by
alarms to their default configuration.
Bypass Alarms
Puts the analyzer into bypass mode, where events
associated with alarms (relays, solenoids, etc) are
ignored. When this button is depressed, the bypass
mode is enabled. When the button is not depressed,
bypass mode is disabled.
Tape Advance
Instructs the analyzer to run the motor to advance the
tape and perform a calibration of the sensor block.
Pressing this button is the equivalent of cycling the
power to the analyzer.
Upgrade
Firmware
Updates the analyzer with the latest firmware. This is
done offline but the user must ensure that the correct
communications port has been selected.
Summary of Contents for ProTech903
Page 2: ......
Page 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Page 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Page 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Page 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Page 190: ...Revision 14 190 August 02 2019...