Chapter 2 Network System Construction
Wired Network System
Chapter 2 Network System Construction
CAUTION
Configure the network as specified by Fukuda Denshi. If the equipment of different network
type is connected, malfunction may occur to the whole network system.
Wired Network System
In this section, connection and setup procedure for wired network is explained.
A wired network system can be constructed by using the LAN cable. Maximum of 48 beds for the DS-LANII
network, maximum of 100 beds for the DS-LANIII network can be connected. The central monitor corresponded to
each wired network is required and the central monitor with the central ID "1" will function as the network
administrator.
DS-LAN Connection
WARNING
Do not connect unspecified equipment to the wired network.
Do not mix devices with DS-LAN II and DS-LAN III setting in the same wired network.The
network may cease and proper monitoring may not be possible.
CAUTION
On a wired network, the alarm generated on the bedside monitor will be output to the
network with a maximum delay of 1 second, and to the central monitor with a total delay of
2.5 seconds.
When connecting to the DS-LAN network, perform "DS-LAN Setup" under [Initial
Settings>System>DS-LAN] and restart the system before connecting the LAN cable.
Use a repeater HUB for DS-LAN II network and a switching HUB for DS-LAN III network.
The default setting of bed ID is "000". If connected to a wired network with the bed ID
unchanged, monitoring on the central monitor will not be possible.
When connecting to a wired network, make sure that there are no other bedside monitors
with the same ID. If there is more than one bedside monitor with the same bed ID, the
duplicated bedside monitors cannot be monitored on the central monitor.
When connecting to the DS-LAN II network, set the ID in the range from 001 to 048. When
connecting to the DS-LAN III network, set the ID in the range from 001 to 100.
If the measurement unit of CO
2
concentration is mmHg and [99 mmHg] is selected for
"Upper Limit of CO
2
(mmHg) Transmission" under [Initial Settings>System>DS-LAN], the
CO
2
value of 100 mmHg or above will be transmitted as 99 mmHg.
NOTE
When the HR/PR numeric data box is set, the alarm limit settings for the following
parameters may not link between the bedside monitor and central monitor. In such case,
perform the setting for each parameter. When changed on the bedside monitor, the alarm
limit settings will link each other.
HR
PR_IBP
Summary of Contents for 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...6 Contents ...
Page 10: ...Contents ...
Page 16: ...Contents ...
Page 46: ...xxx Safety Electromagnetic Compatibility ...
Page 48: ...Chapter 1 Installation of the Unit Contents ...
Page 70: ...Chapter 2 Network System Construction Contents ...
Page 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Page 84: ...Chapter 3 Using the CF card Contents ...
Page 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Page 92: ...Chapter 4 Connection to the External Devices Contents ...
Page 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Page 126: ...Chapter 5 Initial Settings Contents ...
Page 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Page 160: ...Chapter 6 Setup Item Default Value Contents ...
Page 200: ...Chapter 7 Replacement Parts Contents ...
Page 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 214: ...Chapter 9 Maintenance Check Contents ...
Page 228: ...Index 2 Index ...
Page 229: ......
Page 230: ......